Especially service-friendly:
Single shafts and seals can be easily removed, and thus can be directly
cleaned or replaced in the laboratory.
The combination of the stepped design of the ejector and spring made of FKM
reduces the effort needed for ejecting the tips.
Manifold with 360° free rotation in eihther direction to offer optimal
working positions.
Cat.No
|
Channels
|
Volume
range
|
Increment
|
Test
volumen
|
Inaccuracy
|
Imprecision
|
Tip Type
|
BOE 9908050
|
8-ch
|
0,5 - 50 µl
|
0,1 µl
|
50 µl
|
±0,80%
|
±0,40%
|
D,E,F
|
25 µl
|
±1,40%
|
±0,80%
|
5 µl
|
±6,00%
|
±3,00%
|
BOE 9908100
|
8-ch
|
10 - 100 µl
|
0,2 µl
|
100 µl
|
±0,80%
|
±0,30%
|
D,E,F
|
50 µl
|
±1,40%
|
±0,60%
|
10 µl
|
±4,00%
|
±2,00%
|
BOE 9908300
|
8-ch
|
30 - 300 µl
|
0,2 µl
|
300 µl
|
±0,60%
|
±0,30%
|
F
|
150 µl
|
±1,20%
|
±0,60%
|
30 µl
|
±3,00%
|
±1,50%
|
BOE 9912050
|
12-ch
|
0,5 - 50 µl
|
0,1 µl
|
50 µl
|
±0,80%
|
±0,40%
|
D,E,F
|
25 µl
|
±1,40%
|
±0,80%
|
5 µl
|
±6,00%
|
±3,00%
|
BOE 9912100
|
12-ch
|
10 - 100 µl
|
0,2 µl
|
100 µl
|
±0,80%
|
±0,30%
|
D,E,F
|
50 µl
|
±1,40%
|
±0,60%
|
10 µl
|
±4,00%
|
±2,00%
|
BOE 9912300
|
12-ch
|
30 - 300 µl
|
0,2 µl
|
300 µl
|
±0,60%
|
±0,30%
|
F
|
150 µl
|
±1,20%
|
±0,60%
|
30 µl
|
±3,00%
|
±1,50%
|
|